電氣安全測試
Voltage
Range (mains voltage) 90 V to 132 V ac rms
180 V to 264 V ac rms
Range (accessible voltage) 0 V to 300 V ac rms
Accuracy ± (2 % of reading +2 LSD)
Voltage tests Mains, accessible, and point-to-point
Earth resistance
Two-terminal mode test
current/range and accuracy
> 200 mA ac 0.0 to 2.0 Ù ± (2 % of reading + 0.015 Ù)
10 A to 25 A ac 0.0 to 0.2 Ù ± (2 % of reading + 0.015 Ù)
0.2 to 2.0 Ù ± (5 % of reading + 0.015 Ù)
Four-terminal mode test
current/range and accuracy
> 200 mA ac 0.0 to 2.0 Ù ± (2 % of reading + 0.005 Ù)
10 A to 25 A ac 0.0 to 0.2 Ù ± (2 % of reading + 0.005 Ù)
0.2 to 2.0 Ù ± (5 % of reading + 0.005 Ù)
Resistance tests Earth resistance and point-to-point
Equipment current
Mode AC rms
Range/accuracy 0 A to 20 A ± 5 % of reading ± (2 counts or .2 A,
whichever is greater)
Duty cycle 15 A to 20 A, 5 min on/5 min off
10 A to 15 A, 7 min on/3 min off
0 A to 10 A continuous
Leakage current
Modes* AC + DC (True-rms)
AC only
DC only
* Modes are available in all leakage tests with the exception of MAP leakages that
are available only in true-rms
Patient load selection
(input impedance)
AAMI ES1-1993 Fig.1
IEC 60601: Fig 15
IEC 61010: Fig. A-1
Crest factor ≤ 3
Ranges 0 μA to 199.9 μA
200 μA to 1999 μA
2 mA to 10 mA
Frequency response/accuracy DC to 1 kHz ± (1 % of reading + 1 μA)
1 kHz to 100 kHz ± (2 % of reading + 1 μA)
100 kHz to 1 MHz ± (5 % of reading + 1 μA)
Leakage tests Earth (ground wire)
Chassis (enclosure)
Patient (lead to ground)
Patient auxiliary (lead to lead)
Mains on applied part (lead isolation)
Direct equipment
Direct applied part
Alternative equipment
Alternative applied part
Accessible
Point to point
Specifications
ESA620 Electrical Safety Analyzer Fluke Biomedical 3
Mains on applied part
test voltage
110 % of mains at 230 V for IEC 60601
100 % of mains for AAMI at 115 V per AAMI
100 % of mains at 230 V per 62353
Differential leakage
Ranges 10 μA to 199 μA
200 μA to 1999 μA
2 mA to 20 mA
Accuracy ± 10 % of reading ± (2 counts or 20 μA, whichever is greater)
Insulation resistance
Ranges/accuracy 0.5 MÙ to 20 MÙ ± (2 % of reading + 2
counts)
20 MÙ to 100 MÙ ± (7.5 % of reading + 2
counts)
Source test voltage 500 V dc
250 V dc
Insulation resistance tests Mains-PE, AP-PE, Mains- PE, Mains-NE (non-earthed accessible
conductive part) and AP- NE (non-earthed accessible conductive
part)
Mamimum load capacitance 1 μF
ECG performance waveforms
Accuracy ± 2 %
± 5 % for amplitude of 2 Hz square wave only, fixed at 1 mV
Lead II configuration
Waveforms Rates
ECG complex (BPM) 30, 60, 120, 180, and 240
Ventricular fibrillation
Square wave (50 % duty cycle) (Hz) 0.125 and 2
Sine wave (Hz) 10, 40, 50, 60, and 100
Triangle wave (Hz) 2
Pulse (63 ms pulse width) 30 and 60
Power ratings
Mains voltage outlet 120 V ac 230 V ac
Mains voltage inlet power range 90 to 132 V ac rms 180 to 264 V ac rms
Maximum current 20 A 16 A
Hz 50 or 60 50 or 60
Physical case
Dimensions (L x W x H) 31 cm x 23 cm x 10 cm (12.2 in x 9 in x 2.9 in)
Weight 4.7 kg (10.25 lb)
Environmental specifications
Operating temperature 10 °C to 40 °C
Storage temperature -20 °C to 60 °C
Operating humidity 10 % to 90 % non-condensing
Altitude To 2,000 meters
General
Warranty Two-year extended warranty*