生理訊號模擬器

日發行醫療器材有限公司

12-Lead ECGRespArrhythmiaIBPTEMPSpO2NIBP

廠牌型號: 
FlukeBiomedical cufflink/index2/217A
購買日期: 
2012
量測精度及限制說明: 

Specifications

Power 120/250 V ac, 50 W average, 100 W peak, 50/60 Hz

Analog outputs

Cuff pressure 0 mmHg to 499.95 mmHg FS ± 1 % of reading, 10 mV/mmHg

Pulse pressure 0 mmHg to 5 mmHg FS ± 1 of reading, 1 V/mmHg

Digital manometer

Pressure 499.75 mmHg (maximum)

Measurement parameters Instantaneous and peak

Displayed graphics

Dynamic real-time NIBP cuff-pressure waveform programmed peripheral pulse and envelope waveforms

Display

Alphanumeric graphic display (LCD)

Alphanumeric mode 8 lines x 40 characters

Graphics mode 64 vertical x 240 horizontal dot matrix, backlight with viewing angle

adjustment

Digital interfaces

RS-232/Serial Bidirectional; downloads cuff measurement data and controls test

features with a compatible computer or via the medTester 5000C

with the medCheck option.

Parallel printer Centronics compatible

Pulse sync 0 V dc to 5 V dc (TTL)

Cuff mandrel

Interlocking plastic blocks Four cuff circumferences, including: 39.5 cm (large adult),

33 cm (adult), 26.6 cm (small adult), and 20 cm (child)

Truncated plastic cylinders Three neonatal cuff circumferences, including: 14 cm, 10 cm, and

7.6 cm

Pop-off valve testing

Automatic test for operation of the monitors relief valve

Measurement parameters Instantaneous and peak pressure

Maximum pressure 499.75 mmHg

System leak testing

Start pressure 499.75 mmHg max

Elapsed time 60 s (fixed)

Leak-rate range 0.25 mmHg/min to 499.75 mmHg/min

Pump 2 liters/minute minimum (free flow)

Accuracy

Dynamic NIBP Response

Repeatability (Systolic/

Diastalic/Mean)

± 1 % of target value

Cuff pressure ± 1 % of reading ± 1 mmHg

Input overpressure limit ± 1500 mmHg

Autosequences

Up to five user-programmable sequences to test NIBP monitors

with a specific series of CuffLink performance tests, including

static pressure test, leak test, and pop-off test

Up to eight adult-neonatal-arrhythmia dynamic NIBP selections,

each of which can be cycled up to 99 times during the sequence

Printable test report

3 Fluke Biomedical CuffLink Non-Invasive Blood Pressure Simulator

Displayed real-time parameters

Instantaneous cuff pressure 0 mmHg to 300 mmHg

Peak cuff pressure 500 mmHg peak

Inflate/deflate time 0.1 s to 999.9 s

Inflate/deflate rate 0.1 mmHg/s to 999.9 mmHg/s

Total measurement time 0 s to 999.9 s max

Selected heart rate 30 BPM, 40 BPM, 60 BPM, 80 BPM, 120 BPM, 160 BPM, 200 BPM,

and 240 BPM. Selected systolic/diastolic target valuesMean Arterial

Pressure (MAP) target value.

Dynamic non-invasive blood pressure

Simulation of a range of normal, hypertensive and hypotensive dynamic noninvasive blood pressures for

typical adult, infant, and neonatal patients. Generation of normal, bradycardia, and tachycardia rhythm

selections with a wide range of user-programmable peripheral pulse amplitudes (weak, normal and

strong). Compatible with oscillometric NIBP devices.

Preprogrammed target value

selections

Adult systolic/diastolic (MAP) (mmHg): 60/30 (40), 80/50 (62),

100/65 (75), 120/80 (90), 150/100 (115), 200/150 (165), and

255/195 (215)

Neonatal/pediatric systolic/

diastolic

Above selections, excluding 255/195 and 200/150

Repeatability ± 1 % of selected target value

Adult arrhythmia selections Baseline NIBP target value (120/80) (NSR)

Atrial fibrillation (A-Fib)

Premature atrial contraction (PAC)

Premature ventricular contraction (PVC)

Missed beat (MB)

Aberrant sinus conduction (AS)

Preprogrammed peripheral

pulse waveforms

Pulse amplitude at MAP: 2 mmHg (typical adult value)

Pulse volume range: 0 ml to 5.1 ml

Pulse rise time: 80 ms (min)

Heart rates (adult and neonate): 30 BPM, 40 BPM, 60 BPM,

80 BPM, 120 BPM, 160 BPM, 200 BPM, and 240 BPM

Heart-rate accuracy: ± 1 % of selected rate

Preprogrammable target value

shifts

Horizontal axis: Preprogrammed target value selections shifted in 1.0

mmHg steps over a maximum range of ± 300 mmHg to increase or

decrease dynamic pressure values Vertical axis: Relative amplitude

shifted in 1 % increments over a maximum range from 0 % to 200 %

to simulate weak, normal, and strong peripheral pulses

General specifications

Operating temperature 15 oC to 35 oC (59 oF to 95 oF)

Storage temperature 0 oC to 50 oC (32 oF to 122 oF)

Dimensions (LxWxH) 38.1 cm x 31.75 cm x 12.7 cm (15 in x 12.5 in x 5 in)

 

Weight 6.82 kg (15 lb)